WHO/BS/2017.2312 ENGLISH ONLY

EXPERT COMMITTEE ON BIOLOGICAL STANDARDIZATION Geneva, 17 to 20 October 2017 WHO International Collaborative Study of the Proposed 2nd International Standard for Parathyroid Hormone 1-34, Recombinant, Human Ben Cowper*, Peter Rigsby, Thomas Dougall and Chris Burns National Institute for Biological Standards and Control, Blanche Lane, South Mimms, Potters Bar, Herts, EN6 3QG, UK NOTE: This document has been prepared for the purpose of inviting comments and suggestions on the proposals contained therein, which will then be considered by the Expert Committee on Biological Standardization (ECBS). Comments MUST be received by 18 September 2017 and should be addressed to the World Health Organization, 1211 Geneva 27, Switzerland, attention: Technologies, Standards and Norms (TSN). Comments may also be submitted electronically to the Responsible Officer: Dr M. Nübling at email: nueblingc@who.int. © World Health Organization 2017

All rights reserved. Publications of the World Health Organization are available on the WHO web site (www.who.int) or can be purchased from WHO Press, World Health Organization, 20 Avenue Appia, 1211 Geneva 27, Switzerland (tel.: +41 22 791 3264; fax: +41 22 791 4857; e-mail: bookorders@who.int). Requests for permission to reproduce or translate WHO publications – whether for sale or for noncommercial distribution – should be addressed to WHO Press through the WHO web site: (http://www.who.int/about/licensing/copyright_form/en/index.html). The designations employed and the presentation of the material in this publication do not imply the expression of any opinion whatsoever on the part of the World Health Organization concerning the legal status of any country, territory, city or area or of its authorities, or concerning the delimitation of its frontiers or boundaries. Dotted lines on maps represent approximate border lines for which there may not yet be full agreement. The mention of specific companies or of certain manufacturers’ products does not imply that they are endorsed or recommended by the World Health Organization in preference to others of a similar nature that are not mentioned. Errors and omissions excepted, the names of proprietary products are distinguished by initial capital letters. All reasonable precautions have been taken by the World Health Organization to verify the information contained in this publication. However, the published material is being distributed without warranty of any kind, either expressed or implied. The responsibility for the interpretation and use of the material lies with the reader. In no event shall the World Health Organization be liable for damages arising from its use. The named authors alone are responsible for the views expressed in this publication.

WHO/BS/2017.2312 Page 2

Summary The World Health Organization (WHO) Expert Committee on Biological Standardization (ECBS) has recognized the need for a replacement International Standard (IS) for parathyroid hormone 1-34, recombinant, human (rhPTH 1-34) in terms of which the content of therapeutic products can be expressed. We report here the calibration of a candidate standard for rhPTH 1-34 by HPLC assay in terms of a primary calibrant, by an international collaborative study carried out by twelve laboratories in eight countries. The mean of the laboratory estimates of the rhPTH 1-34 content of the candidate standard, coded 15/304, is 0.914 milligrams per ampoule (CV = 2.16%, n=28, expanded uncertainty of 0.902, 0.926 mg/amp) in terms of the primary calibrant. The study also included an assessment of the purity, bioactivity and stability of the candidate standard. The mean of the laboratory estimates of the rhPTH 1-34 purity of the candidate standard is 99.08% (CV = 0.59%, n=11). Bioassay data shows good agreement between the potencies of the candidate standard and the 1st WHO IS, 04/200. The results of this study also indicate that the candidate standard is sufficiently stable to serve as an IS. Therefore, it is proposed that the candidate preparation in ampoules, coded 15/304, is established as the 2nd WHO IS for parathyroid hormone 1-34, recombinant, human with an assigned content of 0.914 mg per ampoule.

Introduction Human PTH 1-34 is the N-terminal, biologically active fragment of parathyroid hormone. The recombinant form of this peptide, expressed in E. coli cells, is commonly known as teriparatide. Teriparatide is currently prescribed in the USA and Europe as a treatment for osteoporosis, under the brand names Forteo and Forsteo respectively (Eli Lilly & Co.). The product is under patent protection until 2019, after which it is anticipated that teriparatide biosimilars will arrive on the market. The 1st IS for parathyroid hormone 1-34, recombinant, human, coded 04/200, was established by the WHO Expert Committee on Biological Standardization (ECBS) in 2007 for the calibration of therapeutic teriparatide (hereafter termed rhPTH 1-34) preparations [1]. This preparation contained rhPTH 1-34 donated by Eli Lilly & Co. Stocks of the 1st IS are low and there is a requirement to replace the standard. The same manufacturer has once again provided a donation of rhPTH 1-34 with which to produce a replacement WHO IS. This material has been formulated and distributed into ampoules, coded 15/304, which have been evaluated in this collaborative study. Participants have been requested to determine the mass content of the candidate standard using a primary calibrant approach. The primary calibrant, coded PRS0404, was previously donated to WHO by Eli Lilly & Co. in 2004, for the purpose of calibrating the 1st IS by international collaborative study. Only half of the donated vials were used, however the purity and content of the remaining vials were recently verified by HPLC analysis, demonstrating that the material is also fit for use as a primary calibrant in this collaborative study.

WHO/BS/2017.2312 Page 3

The aims of this study were therefore: 1. To calibrate the candidate standard, 15/304, in terms of a primary calibrant, PRS0404, by HPLC assay. 2. To determine the purity of the candidate standard, 15/304. 3. To confirm the bioactivity of the candidate standard, 15/304, and its continuity with the 1st WHO IS, 04/200. 4. To determine the stability of the candidate standard, 15/304, by comparison with ampoules stored at elevated temperatures as part of an accelerated degradation stability study.

Participants Twelve laboratories in eight countries took part in the study and are listed alphabetically, by country, in Table 1. Throughout the study, each participating laboratory is referred to by a code number. The code numbers were randomly assigned and do not reflect the order of listing.

Table 1: List of participants in order of country AUSTRALIA Dr Kevin Grant, Dr Tursun Kerim, Dr Sadiqur Talukder, Samuel Ling & Margeret Butt Therapeutic Goods Administration, Biochemistry Section LB, 136 Narrabundah Lane, Symonston ACT 2609 Professor T. John Martin & Patricio Ho St Vincents Institute of Medical Research, 9 Princes Street, Fitzroy VIC 3065 Dr Rolf Humer, Dr Constance Voss, Dr Maria Messeritsch & Dr Bruce Wang Boehringer Ingelheim RCV & GmbH & Co KG, Dr Boehringer Gasse 5-11, 1121 Wien Dr Simon Sauvé & Barry Lorbetskie Health Canada, 251 Sir Frederick Banting Driveway, A/L 2201E, Tunney’s Pasture, Ottawa, Ontario, K1A 0K9 Dr Francois Cano, Dr Francis Poisson & Dr Olivier Garinot ANSM, 143-147 Boulevard Anatole France, 93200 Saint-Denis Dr Jean-Bernard Graff, Sophie Rhinn, Kim Dancheck Lilly France, 2 Rue De Colonel Lilly, 67640 Fegersheim Dr Jai Prakash, Dr M. Kalaivani, Dr Gaurav Pratap Singh & Shruti Rastogi Indian Pharmacopoeia Commission, Biologics Section, Sector 23, Raj Nagar, Ghaziabad, Uttar Pradesh - 201 002

AUSTRALIA AUSTRIA

CANADA

FRANCE FRANCE INDIA

WHO/BS/2017.2312 Page 4 INDIA Dr Renu Jain, Dr Charu Mehra Kamal Dr Meena Kumari & Chanda Sinha National Institute of Biologicals, Plot No. A-32, Sector-62, Institutional Area, Noida - 201 309 Dr Yukari Nakagawa, Chikako Yomota, Akiko Ebisawa & Tomoko Morioka Pharmaceutical and Medical Device Regulatory Science Society of Japan (PMRJ), 2-1-2, Hiranomachi, Chuo-ku, Osaka 541-0046 Dr Ben Cowper National Institute for Biological Standards and Control, Blanche Lane, South Mimms, Potters Bar, Herts, EN6 3QG Dr Paul Matejtschuk & Chinwe Duru National Institute for Biological Standards and Control, Blanche Lane, South Mimms, Potters Bar, Herts, EN6 3QG Dr Bassem Azizeh Bachem, 3132 Kashiwa Street, Torrence, CA 90505

JAPAN

UK

UK

USA

Bulk materials and processing A bulk preparation of highly purified lyophilized rhPTH 1-34 was generously donated to the WHO by Eli Lily & Co. Approximately 8 g of rhPTH 1-34 was dissolved in 8 L of a solution containing 10 mg/ml trehalose and 1 mM acetic acid to give a solution containing approximately 1 mg/ml rhPTH 1-34. The solution was distributed into ampoules as 1.0 ml aliquots. The ampoule contents were freeze-dried and sealed under nitrogen. Ampoules were stored at -20°C.

Characterization of the freeze dried product A total of 7,527 ampoules were produced. Check-weights measured during filling demonstrated a mean fill weight of 1.0066 g (CV = 0.248%, n = 277). The mean residual moisture was estimated to be 0.515% (n = 12) and the mean oxygen head space was estimated to be 0.4% (n =12). No microbial contamination was detected.

Collaborative study for the calibration of 15/304 The collaborative study was organised by NIBSC. The materials provided to participants are summarized in Table 2. A study protocol, shown in Appendix 1, and instructions for use were provided with the samples.

WHO/BS/2017.2312 Page 5

Table 2: Ampouled preparations provided to participants in collaborative study. rhPTH 1-34 preparation Primary calibrant (PRS0404) Candidate 2nd IS for rhPTH 1-34, recombinant, human (15/304) stored at -20°C 1st IS for rhPTH 1-34, recombinant, human (04/200) stored at -20°C Accelerated thermal degradation (ATD) samples of 15/304, stored at +4°C, +20°C, +37°C and +45°C for 179 days (randomly coded A-D) Vial content 1 mg rhPTH 1-34 Nominally 1 mg rhPTH 1-34, plus 10 mg trehalose 0.89 mg rhPTH 1-34, plus 10 mg trehalose

Contents assumed to be identical to 15/304

Participants were asked to provide estimates of the rhPTH 1-34 content of the provided preparations by comparison with primary calibrant via HPLC assay. It was requested that three independent assays be performed, using fresh ampoules/vials for each assay. Due to limited supplies, the inclusion of the 1st IS and ATD samples in only one assay was permitted. Participants were also requested to provide estimates of the rhPTH 1-34 purity of the candidate standard via HPLC. Additionally, in order to provide supporting information for the use of the candidate standard in a bioassay, participants were requested to carry out the bioassay normally in use in their laboratory. It was requested that three independent assays be performed, using fresh ampoules/vials for each assay, and that each assay include candidate standard, 15/304, and 1st IS, 04/200, preparations at no less than five dilutions in the linear part of their dose-response curve. In instances where there was not a fresh ampoule for subsequent assays, it was suggested that fresh dilutions be made from frozen stock solutions. The assays provided by each laboratory are shown in Table 3.

Table 3: Methods contributed by each laboratory Lab 1 2 3 4 Method HPLC (content and purity) HPLC (content and purity) HPLC (content and purity) In vitro Bioassay As protocol. As protocol. Mobile phase composition adjusted to obtain desired retention times in both tests. 100 Å column pore size. As protocol. cAMP measurement in UMR106 cells, stably transfected with a cAMP Responsive Element (CRE)-Luciferase gene construct. Detection via luminescence. Comments

WHO/BS/2017.2312 Page 6 5 6 7 8 9 10 11 12 HPLC (content and purity) HPLC (content) In vitro Bioassay HPLC (content and purity) HPLC (content and purity) HPLC (content and purity) HPLC (content and purity) HPLC (content and purity) As protocol. As protocol. cAMP measurement in UMR106.01 cells. Detection via radioimmunoassay. As protocol. Mobile phase composition adjusted to obtain desired retention times in purity test. As protocol. 100 Å column pore size. As protocol. Mobile phase composition adjusted to obtain desired retention times in both tests. As protocol. Mobile phase composition adjusted to obtain desired retention times in both tests. As protocol. Mobile phase composition adjusted to obtain desired retention times in both tests.

Results Data returned for analysis Data were contributed by 12 laboratories who performed in total 30 individual HPLC content assays, 11 HPLC purity assays and 9 In vitro bioassays.

Assay validity The majority of HPLC content and purity assays satisfied the system suitability criteria, thus allowing valid estimates of rhPTH 1-34 content and purity to be calculated. However Assay 2 for content from Laboratory 6 did not satisfy the system suitability criteria for rhPTH 1-34 peak RSD% or peak tailing, and was therefore excluded in the analysis. Assay 3 for content from Laboratory 6 was also excluded as an outlier.

Estimated content of the candidate standard, 15/304, in terms of the primary calibrant Estimates of rhPTH 1-34 content (mg/amp) reported by each laboratory were checked and verified. In the results sheet originally provided to participants, rhPTH 1-34 content (mg/ampoule) was calculated by multiplying the estimated rhPTH 1-34 concentration (mg/ml) by 4 (the ampoule reconstitution volume, in ml). Participants were also requested to report the accurately weighed mass of solvent added to ampoules during reconstitution. Following data receipt, rhPTH 1-34 content estimates were re-calculated, by instead multiplying the estimated rhPTH 1-34 concentration by the mass of solvent added to the ampoule. During the collaborative study of the 1st WHO IS, 04/200, the primary calibrant, PRS0404, was estimated to contain 0.997 mg/vial rhPTH 1-34. PRS0404 was then provided to participants with an assigned content of 1.0 mg/vial and a correction was later applied to the mean of all estimates

WHO/BS/2017.2312 Page 7 of rhPTH 1-34 content. The same approach has been taken here for the candidate standard, 15/304. Estimates for the rhPTH 1-34 content of the candidate standard, 15/304, in terms of the primary calibrant, PRS0404, are summarized in Table 4. The estimated mean content across nine laboratories is 0.917 mg/amp, which corrected for the estimated content of the primary calibrant (0.997 mg/vial) is 0.914 mg/amp. The relative standard uncertainty was determined as shown in Table 5 with an overall estimate of 0.65%. A final estimate of content for the candidate standard, 15/304, is 0.914 (0.902, 0.926) mg/amp (expanded uncertainty with coverage factor of k=2 taken to correspond to a 95% level of confidence). The rhPTH 1-34 content of the candidate standard, 15/304, was also estimated in terms of the 1st IS, using its assigned content of 0.89 mg/amp, and found to be 0.936 mg/amp (95% confidence interval of 0.931, 0.942), as summarized in Table 6.

Estimated content of the 1st WHO IS, 04/200, in terms of the primary calibrant Estimates of the rhPTH 1-34 content of the 1st WHO IS, 04/200, in terms of the primary calibrant, PRS0404, are summarized in Table 6. The estimated mean content across nine laboratories is 0.871 (0.860, 0.883) mg/amp (expanded uncertainty calculated the same way as for the candidate standard), or 0.869 (0.857, 0.880) mg/amp when correcting for primary calibrant content. The collaborative study of the 1st IS, carried out in 2007, assigned an estimated content of 0.89 mg/amp. Although an uncertainty was not assigned to the 1st IS at the time, this has been retrospectively determined to be 0.868, 0.912 mg/amp.

Assessment of 15/304 purity Estimates reported by each laboratory were checked and verified prior to value assignment. Where no peak threshold had been employed, a relative peak area (%RPA) threshold of ≥0.03% was applied. Estimates for the rhPTH 1-34 purity of candidate standard, 15/304, are summarized in Table 7. The mean estimated purity is 99.08% (95% confidence interval of 98.69, 99.47).

Assessment of stability The relative contents of the accelerated thermal degradation samples were used to fit an Arrhenius equation relating degradation rate to absolute temperature assuming first-order decay [2] and hence predict the degradation rates when stored at -20°C. Estimates of the content of candidate standard, 15/304, ampoules stored at elevated temperatures for a period of 179 days, in terms of the primary calibrant, PRS0404, are summarized in Table 8. Estimates of the relative potencies of these ampoules, in terms of 15/304 ampoules stored at 20oC, are summarized in Table 9. Laboratory 10’s reported data was excluded from the analysis due to having outlying results. Analysis showed a predicted loss of content per year of 0.042% when stored at -20°C.

Bioassay data Two laboratories each returned data from bioassays including the current standard, 04/200, and the candidate standard, 15/304. Bioassay data was analysed in CombiStats v5.0 [3] using a sigmoid curves model and checked for significant non-parallelism between samples using

WHO/BS/2017.2312 Page 8 analysis of variance. All assays showed no deviation from parallelism when considered significant at the 1% level (p<0.01). Relative potency estimates are shown in Table 10.

Discussion & conclusions The aim of this collaborative study was to assign a value to the candidate 2nd WHO IS for rhPTH 1-34, 15/304, using a HPLC assay. Estimates for the content of 15/304 were in good agreement and the content in terms of the primary calibrant, PRS0404, was determined as 0.914 mg per ampoule (CV 2.16%; n=28) (Table 4). The value includes a correction for the estimated content of the primary calibrant, PRS0404, which was provided to participants with an assigned value of 1.0 mg/vial for HPLC analysis, but was accurately estimated to contain 0.997 mg/vial during the collaborative study of the 1st WHO IS for rhPTH 1-34, 04/200. The relative uncertainty of the candidate standard has been estimated as 0.65% (Table 5) giving rise to an expanded uncertainty of 0.902, 0.926 mg/amp. The rhPTH 1-34 content of the candidate standard, 15/304, has also been estimated in terms of the 1st IS, 04/200, using its assigned content of 0.89 mg (Table 6). The resulting estimate of 0.936 mg/amp differs by 0.022 mg (2.4%) from the estimate of 0.914 mg/amp derived in terms of the primary calibrant, which if not accounted for could cause a shift in potency assignments of rhPTH 1-34 therapeutic products upon uptake of the replacement IS. The difference is likely to be caused by inaccurate content assignment of the 1st IS. Repeat analysis of the 1st IS in this study estimates an rhPTH 1-34 content of 0.869 mg/amp in terms of the primary calibrant, which is lower (by 2.36%) than its assigned content of 0.89 mg/amp. It should be noted that the collaborative study of the 1st WHO IS reported significant deviations in estimates between laboratories performing HPLC assays (CV 5.05%, n=40), and the expanded uncertainty has been retrospectively determined here to be 0.868, 0.912 mg/amp. In this study the estimates for the 1st IS are in much closer agreement (CV 1.46%, n=10) with a reduced uncertainty of 0.857, 0.880 mg/amp. Therefore it is believed that the value of 0.869 mg/amp reported here is a more accurate estimation of rhPTH 1-34 content in the 1st IS. The estimated rhPTH 1-34 content of the candidate standard, 15/304, relative to the 1st IS, 04/200, using a corrected content of 0.869 mg, is 0.914 mg/amp; equal to the estimate relative to the primary calibrant. The study also aimed to evaluate the purity of the candidate standard via HPLC, giving rise to an estimated mean rhPTH 1-34 purity of 99.08% (95% confidence interval of 98.69, 99.47) (Table 7). The candidate standard is therefore of very high purity, of a similar order to the current IS, and is therefore suitable to serve as a replacement IS. The bioactivity of the candidate standard has also been demonstrated through bioassays performed in two laboratories. Data from laboratory 4 demonstrates very close agreement between the potencies of the candidate standard and the 1st WHO IS (Table 9). Whilst data from laboratory 7 showed more variability in relative potency, taken together the data demonstrates that use of the candidate standard would ensure continuity of rhPTH 1-34 measurements in bioassays. Finally, the stability of the candidate standard, 15/304, was assessed by the measurement of accelerated thermal degradation (ATD) samples via HPLC assay (Table 8). The candidate standard

WHO/BS/2017.2312 Page 9 was predicted to exhibit a loss of content of 0.042% per year when stored at -20°C. This value is higher than expected, especially when compared with data from the collaborative study of 1st WHO IS for rhPTH 1-34, which predicted an annual loss of 0.001% based on HPLC assays. The candidate standard, 15/304, contains the same active ingredient from the same source, formulated in the same manner as the 1st IS. Mean moisture levels are lower in the candidate standard (0.5%) than the 1st WHO IS (1.31%), and oxygen levels, although not recorded for the 1st IS, are low in the candidate standard (0.4%). Therefore the candidate standard would be expected to be at least as stable as the 1st IS, for which HPLC chromatograms have been acquired in this study and exhibit little evidence of degradation. Furthermore closer inspection of the relative potencies of ATD samples reveals that laboratory 2 observed a larger decrease in relative potency at +4oC than all other laboratories (Table 9). Exclusion of laboratory 2 from the analysis reduces the predicted annual loss to 0.005% per year at -20oC. Considering all of the above discussion, it is anticipated that the candidate standard, 15/304, is sufficiently stable to serve as a WHO IS stored at -20oC. However ATD samples will be continually monitored at NIBSC over the next 1-2 years, and if further analysis leads to predictions of significant losses then the material will be transferred to -70oC storage. Prior to the establishment of the 1st WHO IS for rhPTH 1-34, 04/200, a NIBSC Research Reagent for PTH 1-34 (82/508) was in use, which was assigned units of activity (355U) relative to a previous in-house preparation (75/596) thought to contain 0.03 mg (i.e. using a conversion factor of 10000 units/mg). Although rhPTH 1-34 therapeutic preparations are routinely administered in mass units, activity units are sometimes quoted in addition, which may originally have been based on the NIBSC Reference Reagent. Consequently the 1st WHO IS was assigned a unitage in addition to its mass content (0.89 mg; 8900IU; 10000 U/mg). Since rhPTH 1-34 International Standards are suitable for use in bioassays and immunoassays, an assigned unitage may continue to be useful, and therefore the assigned mass content of the candidate standard, 15/304, should also be converted to units (0.914 mg; 9140IU; 10000 U/mg).

Proposal It is recommended that the preparation in ampoules coded 15/304 is established as the 2nd IS for parathyroid hormone 1-34, recombinant, human, with an assigned content of 0.914 mg/amp. To maintain continuity with the assigned unitage of the 1st WHO IS for rhPTH 1-34, it is also proposed that this material be assigned an ampoule content of 9140 International Units (1 IU approximately equivalent to 100 ng).

Acknowledgements We gratefully acknowledge the important contributions of all the participants in the collaborative study, Eli Lilly & Co., who donated the bulk material, and the Centre for Biological Reference Materials, NIBSC for preparation and despatch of the ampouled materials.

WHO/BS/2017.2312 Page 10

References [1] WHO International Collaborative Study of the Proposed 1st International Standard for Parathyroid Hormone 1-34, Recombinant, Human. Geneva. World Health Organisation, Expert Committee on Biological Standardisation. WHO/BS07.2063 [2] Kirkwood, T. B. (1977). Predicting the stability of biological standards and products. Biometrics 33(4): 736-742. [3] Daas A. Combistats. 5.0 edn: EDQM - Council of Europe, 2013.

WHO/BS/2017.2312 Page 11

Table 4: Laboratory estimates of rhPTH 1-34 content in the candidate standard, 15/304, calculated relative to the primary calibrant, PRS0404, with an assumed content of 1 mg/vial Lab 01 02 03 05 06 08 09 10 11 12 Assay 1 0.915 0.930 0.911 0.924 0.884 0.923 0.958 0.900 0.939 0.919 Assay 3 0.920 0.928 0.906 0.936 1.374# 0.915 0.914 0.934 0.931 0.905 Mean Standard Uncertainty Expanded Uncertainty (k = 2) Assay 2 0.911 0.927 0.916 0.922 0.979* 0.924 0.896 0.850 0.936 0.911 Mean 0.915 0.928 0.911 0.927 0.884 0.921 0.923 0.895 0.935 0.912 0.917 0.65% (0.905, 0.929) 0.914 (0.902, 0.926)

See Table 5

Corrected Mean† Corrected Expanded Uncertainty (k = 2)†

* excluded as assay did not meet system suitability criteria # excluded as an outlying result † corrected for estimated content of primary calibrant (0.997 mg/vial)

Table 5. Uncertainty of measurement determination for 15/304 and 04/200 Relative Standard Uncertainty 0.04% 0.40% 0.25% 0.45% 0.65%

Variability Source Between Laboratory Between Assay Homogeneity of Fill Primary Calibrant

Estimation Method Using a random effects model treating ‘lab’ as a random effect %CV of fill weights for 277 ampoules From previous international collaborative study data Combined relative standard uncertainty

WHO/BS/2017.2312 Page 12

Table 6: Laboratory estimates of rhPTH 1-34 content in the candidate standard, 15/304, calculated relative to the 1st IS, 04/200, with its assigned content of 0.89 mg and corrected content of 0.869 mg/amp, and in both 15/304 and 04/200 calculated relative to the primary calibrant, PRS0404, with an assumed content of 1 mg/vial Lab 01 02 03 05 06 08 09 10 11 12 Mean 95% LCL 95% UCL 15/304 vs PRS0404 0.915 0.928 0.911 0.927 0.884 0.921 0.923 0.895 0.927 0.912 0.917 (0.914†) 0.905 (0.902†)* 0.929 (0.926†)* 04/200 vs PRS0404 0.866 0.875 0.875 0.877 0.853 0.875 0.894 0.862 0.882 0.854 0.871 (0.869†) 0.862 (0.857†)* 0.880 (0.880†)* 15/304 vs 04/200 (0.89 mg/amp) 0.936 0.948 0.931 0.938 0.921 0.938 0.932 0.933 0.942 0.944 0.936 0.931 0.942 15/304 vs 04/200 (0.869 mg/amp) 0.914 0.926 0.909 0.916 0.899 0.915 0.910 0.911 0.920 0.922 0.914 0.903 0.925

*expanded uncertainty with coverage factor k=2 taken to correspond to a 95% level of confidence † corrected for estimated content of primary calibrant (0.997 mg/vial)

WHO/BS/2017.2312 Page 13

Table 7: Laboratory estimates of rhPTH 1-34 purity in the candidate standard, 15/304 Lab 01 02 03 05 08 09 10 Injection 1 99.61 98.81 99.09 99.01 99.90 99.85 98.42 98.25 98.30 99.29 99.24 Injection 2 99.62 98.78 99.14 98.95 99.90 99.86 98.34 98.30 98.36 99.38 99.24 Injection 3 99.61 98.76 99.09 98.90 99.90 99.84 98.39 98.27 98.46 99.42 99.23 Mean 95% LCL 95% UCL Mean 99.61 98.78 99.10 98.95 99.90 99.85 98.39 98.27 98.37 99.36 99.24 99.08 98.69 99.47

11 12

* excluded as data did not meet system suitability criteria

Table 8: Laboratory estimates of rhPTH 1-34 content of accelerated degradation study samples of 15/304 (179 days incubation), calculated relative to the primary calibrant, PRS0404, with an assumed content of 1 mg/vial Lab 01 02 03 05 06 08 09 10* 11 12 Mean 95% LCL 95% UCL 15/304 (-20°C) 0.911 0.930 0.911 0.924 0.884 0.923 0.914 0.900* 0.936 0.919 0.920 0.911 0.928 B (+4°C) 0.913 0.910 0.912 0.915 0.886 0.922 0.914 0.804* 0.932 0.897 0.914 0.906 0.921 A (+20°C) 0.913 0.910 0.912 0.927 0.888 0.917 0.905 0.844* 0.931 0.900 0.917 0.907 0.926 C (+37°C) 0.899 0.896 0.897 0.912 0.909 0.907 0.892 1.093* 0.918 0.884 0.901 0.892 0.910 D (+45°C) 0.888 0.892 0.889 0.901 0.886 0.897 0.892 0.879* 0.901 0.871 0.892 0.884 0.900

* excluded as outlying results

WHO/BS/2017.2312 Page 14

Table 9: Laboratory estimates of rhPTH 1-34 relative potencies of accelerated degradation study samples of 15/304 (179 days incubation), calculated relative to samples stored -20oC. Lab 01 02 03 05 06 08 09 10 11 12 B (+4°C) 1.00 0.98 1.00 0.99 1.00 1.00 1.00 0.89* 1.00 0.99 A (+20°C) 1.00 0.98 1.00 1.00 1.00 0.99 0.99 0.94* 0.99 0.99 C (+37°C) 0.99 0.96 0.98 0.99 1.03 0.98 0.98 1.21* 0.98 0.98 D (+45°C) 0.97 0.96 0.98 0.98 1.00 0.97 0.98 0.98* 0.96 0.96

* excluded as outlying results

Table 10: Estimated potencies of candidate standard, 15/304, relative to the current standard, 04/200, in bioassays Lab Assay 1 2 3 4 5 6 1 2 3 95% LCL 0.91 0.88 0.84 0.85 0.73 0.88 0.59 0.64 0.90 Estimate 0.99 0.98 0.94 0.98 0.84 1.01 0.79 0.71 1.11 95% UCL 1.07 1.10 1.05 1.14 0.97 1.17 1.05 0.79 1.37

04

07

WHO/BS/2017.2312 Page 15

Appendix 1 COLLABORATIVE STUDY PROTOCOL - ESTABLISHMENT OF THE 2nd WHO INTERNATIONAL STANDARD FOR PARATHYROID HORMONE 1-34, RECOMBINANT, HUMAN (15/304) INTRODUCTION PTH 1-34 is the N-terminal, biologically active fragment of human parathyroid hormone. It is prescribed, under the brand name Forteo, in the USA and Europe as a treatment for osteoporosis. In 2004 the World Health Organisation (WHO) Expert Committee on Biological Standardisation (ECBS) recognised the need for an International Standard (IS) for recombinant, human, parathyroid hormone 1-34 (rhPTH 1-34, also known as teriparatide) in terms of which the content of therapeutic products can be expressed. This resulted in the establishment of the 1st WHO IS for rhPTH 1-34 at NIBSC (Code No. 04/200) in 2007. Stocks of this standard are expected to be exhausted within the next two years, necessitating the development of a replacement 2nd IS. A preparation of rhPTH 1-34 has been received at NIBSC and distributed into ampoules (Code No. 15/304) as a candidate 2nd WHO IS, following procedures recommended by WHO (1). It is proposed to carry out an international collaborative study to establish this replacement standard, by calibration in terms of a primary calibrant of rhPTH 1-34 using HPLC, and confirmation of bioactivity via bioassay. The primary calibrant (Code No. PRS0404) was previously donated for the purpose of calibrating the 1st IS by international collaborative study, and assigned a mass value of 1.0 mg per vial by amino-acid analysis and UV spectroscopy. Only half of the donated vials were used, and the remainder were stored at -80oC. The purity and content of these vials has recently been verified by HPLC analysis at NIBSC, demonstrating that the material is fit for use as a primary calibrant in this collaborative study. Therefore, the aim of the study is to calibrate the candidate standard (15/304) in terms of the primary calibrant (PRS0404) by HPLC, and to establish continuity with the existing International Standard for rhPTH 1-34 (04/200).

MATERIALS Preparations supplied to participants in collaborative study. The materials for this study are listed in Table 1. The study includes accelerated thermal degradation (ATD) samples of the candidate standard, which are coded by letter (A-D) in random order. Triplicate primary calibrant (PRS0404) vials and multiple candidate standard (15/304) ampoules are provided for analysis. Due to limited availability, restricted numbers of 1st WHO IS (04/200) and 15/304 ATD samples (A-D) are provided.

WHO/BS/2017.2312 Page 16 rhPTH 1-34 preparation Primary calibrant (PRS0404) Candidate standard (15/304), stored at -20 °C 1st WHO IS (04/200) Accelerated thermal degradation (ATD) samples of the candidate standard, stored at +4°C, +20°C, +37°C and +45°C (randomly coded A-D) Vial content 1 mg rhPTH 1-34 Nominally 1 mg rhPTH 1-34, plus 10 mg trehalose 0.89 mg rhPTH 1-34, plus 10 mg trehalose Content assumed to be identical 15/304

Table 1. Materials provided to collaborative study participants.

Handling of material On receipt all vials and ampoules should be stored at -20°C until use. Before opening, vials and ampoules should be brought to room temperature to minimise moisture uptake. Bulk material and preparation of rhPTH 1-34 ampoules Highly purified rhPTH 1-34, expressed in E. coli cells, was kindly donated by Eli Lilly & Co. (Indianapolis, IN, USA). The preparation was received as a lyophilised white powder, which was formulated at 1 mg/mL with pre-filtered trehalose (10 mg/mL) in 1 mM acetic acid and dispensed into 5 mL ampoules at 1 mL per ampoule. Ampoules were then lyophilised and sealed. A total of 7,527 ampoules were produced, with a mean fill mass of 1.0066 g, a mean dry weight of 0.0106 g, and a mean residual moisture content of 0.5154%. No microbial contamination was detected.

TESTS REQUESTED HPLC for content assay and purity evaluation Participants are asked to provide estimates of the rhPTH 1-34 content of provided preparations, following the enclosed HPLC protocol (Appendix A) as closely as is practicable. It is requested that participants perform three independent assays, and that the rhPTH 1-34 content of the candidate standard (15/304), its ATD samples (A-D), and the 1st WHO IS (04/200) is estimated by comparison with the primary calibrant (PRS0404) using its assigned mass value of 1.0 mg per vial. Because restricted numbers of 1st WHO IS (04/200) and ATD samples (A-D) have been provided, these may be included in only one HPLC assay. Participants are also asked to provide estimates of the purity of the candidate standard (15/304) following the enclosed HPLC protocol (Appendix B) as closely as is practicable. Bioassay In order to provide supporting information for the use of the candidate standard (15/304) in a bioassay participants are requested to carry out the bioassay normally in use in their laboratory. Where possible it is requested that at least two independent assays be performed, using fresh ampoules, and that each assay includes candidate standard (15/304) and 1st WHO IS (04/200) preparations at preferably no less than five dose levels in the linear part of the dose-response

WHO/BS/2017.2312 Page 17 curve. In instances where there is not a fresh ampoule for subsequent assays, it is suggested that fresh dilutions are made from frozen stock solutions. It is recommended that the contents of each ampoule are reconstituted in a suitable assay diluent e.g. PBS or saline (not water) and appropriate dilutions made from this stock solution. Since there are likely to be extensive dilutions to achieve required assay doses, protein cover (typically 0.1% BSA or HSA) to prevent surface adsorption should be provided. There is some evidence that if the initial dilution is carried out in a relatively large volume, variation resulting from the numerous dilution steps is minimized. If practicable, an initial dilution volume of 5 ml for all preparations is therefore suggested.

REPORT A preliminary report will be prepared and circulated to all participants for comment before submission to the Expert Committee on Biological Standardization of WHO. In the report, participating laboratories will be identified by a laboratory number only and any requests to treat information in confidence will be respected.

REFERENCES 1. WHO Tech Rep Ser No 800, 1990, 181-214

For further information, please contact: Ben Cowper National Institute for Biological Standards and Control Blanche Lane South Mimms Hertfordshire EN6 3QG UK Tel: 44 (0) 1707 641242 e-mail: ben.cowper@nibsc.org http://www.nibsc.ac.uk

WHO/BS/2017.2312 Page 18

APPENDIX A HPLC ASSAY OF THE CANDIDATE STANDARD (15/304) USING THE PRIMARY CALIBRANT (PRS0404) AS A REFERENCE STANDARD 1. KEY EQUIPMENT & MATERIALS  A gradient high performance liquid chromatography (HPLC) instrument equipped with a variable wavelength UV detector, loop injection valve, a column oven, and a refrigerated auto-sampler. HPLC column

Zorbax 300SB-C18 (4.6 x 150 mm, particle size 3.5 µm, pore size 300 Å) or validated equivalent.  0.2 M sulphate buffer, pH 2.3 E.g. Dissolve 28.4 g of sodium sulphate (e.g. ACS reagent, ≥99%, Sigma -Aldrich Cat. No. 239313) in 900 ml water. Adjust the pH to 2.3 with 85% phosphoric acid (85-90%, for HPLC, Sigma-Aldrich Cat. no. 79606) and make up to 1 L with water. Mobile phase A 90% sulphate buffer, 10% acetonitrile. Mix 1 volume of acetonitrile (e.g. HiPerSolv CHROMANORM® for HPLC, VWR Cat. No. 83639.320) with 9 volumes of sulphate buffer. Mobile phase B 50% sulphate buffer, 50% acetonitrile. Mix equal volumes of acetonitrile and sulphate buffer. If necessary, incubate at 37oC for 10-15 minutes to ensure complete dissolution of reagents.

NB. All aqueous buffers should be prepared in double distilled, or HPLC grade water (e.g. Chromasolv® water for HPLC , Sigma-Aldrich, Cat. No. 270733), filtered and degassed.

2. STANDARD AND SAMPLE PREPARATION  Perform three independent HPLC assays, using fresh ampoules/vials for each assay. Due to limited ampoule availability, the 1st WHO IS (04/200), and coded ATD samples (A-D) may be included in just one assay. Preparation of the standard curve of the primary calibrant (PRS0404). A standard curve of the mean peak area resulting from a minimum of five quantities of rhPTH 134 should be prepared. Dilutions should be prepared in duplicate, using mobile phase A as the diluent. The precise mass of solvent added to the primary calibrant vial during the initial reconstitution step should be measured using a weighing balance, and reported in the provided results sheet, alongside details of serial dilutions. For example, reconstitute the primary calibrant vial with 2 mL mobile phase A (and record the weight of solvent added to the vial), to obtain a rhPTH 1-34 concentration of 0.5 mg/mL. Use this

WHO/BS/2017.2312 Page 19 stock solution to accurately prepare dilutions (in duplicate) of 0.35, 0.2, 0.1 and 0.05 mg/mL. During HPLC analysis 20 µl injections of each preparation will therefore provide a standard curve with determinations of 10, 7, 4, 2 and 1 µg rhPTH 1-34.  Preparation of ampoules; 04/200, 15/304, A, B, C, D Reconstitute ampoule contents in 4 ml mobile phase A to obtain a rhPTH 1-34 concentration of 0.25 mg/mL. The precise mass of solvent added to each ampoule during reconstitution should be measured using a weighing balance, and reported in the provided results sheet.

NB. Samples are stable for 48 hours when stored at refrigerated temperature in a properly sealed container.

3. SAMPLE ANALYSIS  HPLC operating conditions:  Flow rate  Injection volume  Column temperature  Autosampler  Run time  Detection  Mobile phase 1.0 mL/min 20 μL 40 °C 5 °C 20 mins UV at 214 nm 63% mobile phase A/37% mobile phase B (isocratic)

If precipitation occurs then apply mild heading (2025 oC). Stir continuously during analysis. NB. The approximate retention time of rhPTH 1-34 is 7-11 mins. The mobile phase composition may be adjusted in order to obtain the desired retention time. Please report any deviations from the stated protocol.  System suitability  Perform three injections of primary calibrant (PRS0404) containing 0.5 mg/mL rhPTH 134. See Figure 1 for a typical chromatogram.  The relative standard deviation (RSD) of the rhPTH 1-34 peak area must be ≤1.25%.  The mean for peak tailing (asymmetry) must be ≤1.5 for the rhPTH 1-34 peak, according to the USP definition: